Made from cellulose nitrate and acetate
Type ME
• Economical
• Ideal for gravimetric analysis
• Specially for aqueous solutions
• Very good for use in contamination tests
• Constant weight
• Hydrophilic
• Can be used up to 125°C
• Can be autoclaved at up to 121°C
• Pore sizes from 0.2 to 3 µm
Product Selection
Type |
Nominal
pore size µm |
Thickness
µm |
Water flow rate
p = 0.9 bar ml/min/cm2 |
Air flow rate
p = 3 mbar ml/min/cm2 |
Bubble
point bar |
ME-24 |
0.2
|
135
|
25
|
–
|
3.7
|
ME-25 |
0.45
|
135
|
45
|
25
|
2.5
|
ME-26 |
0.6
|
135
|
110
|
45
|
1.5
|
ME-27 |
0.8
|
140
|
170
|
80
|
1.3
|
ME-28 |
1.2
|
140
|
240
|
100
|
0.8
|
ME-29 |
3
|
150
|
400
|
140
|
0.7
|